JUN-AIR Dental Compressed Air System User Guide CI-101A (Rev B)
Problem Possible Cause(s) Possible Solution(s)
5. Compressor cycles when there is
no air demand from the operator/
operatories
a. Leak in the compressor system a. See 4C
b. Leak in the ofce air system b. Look at the moisture indicator (see KEY PARTS
to locate.) If it is blue....With the power switch
ON, (if necessary) drain the storage tank to 80
psi to start the compression cycle. When the
cycle shuts off at 100 psi, close the air outlet
valve. Wait 5 minutes. Open the air outlet valve.
If the pressure drops, (by more than 5 psi) the
air leak is in the office air system or delivery units
and not in the Compressor System. Call your
authorized dealer for service.
NOTE: If it is pink - see #6 below.
6. Moisture indicator is pink* a. Leak in the ofce air system a. If the moisture indicator is pink, there is too much
moisture in the system.
CAUTION: Call your authorized dealer for
service.
b. Compressor keeps cycling b. Check the “SPECIFICATIONS”. There may be
excessive demands placed on the Compres-
sor System. A larger capacity model may be
required.
c. The solenoid valve requires attention c. It is normal to hear air escaping from the sole-
noid valve when the compressor shuts off. If air
does not escape, the valve could be clogged or
sticking. Call your authorized dealer for service.
* In initial installations, the moisture indicator may appear pink before operating the system, this is due to ambient air in the sys-
tem and will change to blue after operation of the system.
TROUBLESHOOTING CHART
DIAGNOSTIC PROCEDURE FOR DEFECTIVE COMPRESSOR(S)
1. Turn compressor power switches to the OFF
position.
WARNING: Always turn off compressor
and remove power from unit when ser-
vicing or removing electrical cover. Lock
out power at the breaker prior to servic-
ing.
2. Reset the circuit breaker if it was previously
tripped.
3. Test compressors by turning ONE on at a
time using compressor switches located in
electrical panel. (Start only 1 compressor at a
time and disconnect power between testing
compressors). If the motor fails to start, or the
circuit breaker trips, the problem may be in that
compressor. Leave the compressor switch for
the defective compressor in the OFF position.
Call your authorized dealer for service.
NOTE: One or two compressors (depending
on system size) may be run temporarily while
waiting for service.
4. If all compressors run independently, but will not
run together, check the line voltage. If the voltage
is within the min/max. voltage required in SITE
REQUIREMENTS, call your authorized dealer for
service.