4
CI-101A (Rev B) JUN-AIR Dental Compressed Air System User Guide
© 2011, JUN-AIR
We reserve the right to make any alterations which may be due to any technical improvements
Printed in the USA
INTENDED USE:
To provide compressed air during general examina-
tions and procedures conducted by qualified dental
professionals.
JUN-AIR compressor systems meet or exceed the
most current and highest safety standards, which
are:
CSA C22.2 601.1-M90 (R2005)
UL60601-1 classied revision 2006/04/26
- Classification: Class I, permanently fixed MAINS
operated equipment
NFPA 99C level 3 gas system requirements
compliance revision 2005
NFPA 70 (National Electric Code) revision 2008
ISO 9001:2000
Ingress protection: IPX0
To ensure the safety potential of this equip-
ment is achieved, please:
Make sure your equipment is installed according to
the instructions provided in this manual and make
sure the installation checklist is completed prior to
starting the equipment.
Equipment is not suitable for use in the presence of
a ammable anesthetic mixture with air/oxygen or
nitrous oxide. DO NOT OPERATE THE EQUIPMENT
IF THESE CONDITIONS EXIST.
Protection against electrical shock:
Provide proper grounding per NFPA 70 (NEC 2008).
Do not become a current path for the equipment to
ground through your body.
Transportation/Storage Conditions:
Temperature range -29 °C / -20 °F to
60 °C / 140 °F. Relative Humidity 10 to 90%
(non-condensing). Barometric pressure minimum of
372 MM Hg (0.49 ATM).
Keep system dry at all times.
Do not stack units during shipment or
installation.
Important: Refer servicing and installation to
an authorized service representative.
IMPORTANT SAFETY INSTRUCTIONS AND REGULATORY INFORMATION
WARNING: The system should only be
installed by qualified personnel. The
system should be installed in a clean,
dry, well ventilated area on a solid, level
surface.